Abstract
The determination of photographic processing solutions through buffer curve analysis was investigated to determine its advantages, limitations, and general usefulness. A microprocessor controlled titration system was set up to perform up to 16 unassisted titrations and record the data on a floppy disk system at a speed five times faster than the speed of manual titrations. The ingredients are not detectable by this analysis if their concentrations are much less than 1/10th the normality of the ingredient with the largest concentration that is detectable by this analysis. There is linearity between the concentration of a single ingredient solution and the volume of titrant required to reach the final endpoint, but this linearity fails when more than one ingredient is in the solution. This analysis is able to detect a significant number of ingredients used in photographic processing solutions. The analysis can be useful in analyzing competitors' products and photographic processor control.
Library of Congress Subject Headings
Photographic sensitometry
Publication Date
5-1-1978
Document Type
Thesis
Department, Program, or Center
School of Photographic Arts and Sciences (CIAS)
Advisor
Francis, Ronald
Recommended Citation
Winslow, Richard, "Determination of unknown photographic processing solutions through buffer curve analysis" (1978). Thesis. Rochester Institute of Technology. Accessed from
https://repository.rit.edu/theses/5151
Campus
RIT – Main Campus
Comments
Note: imported from RIT’s Digital Media Library running on DSpace to RIT Scholar Works. Physical copy available through RIT's The Wallace Library at: TR280 .W55